Error ID: 25226
The autosampler failed to complete injection sequence correctly. The autosampler can move and draw sample from the draw position and generates the error message.
Probable cause | Suggested actions | |
---|---|---|
1 | Sample container is not installed correctly in the pallet. | Check if the sample container is installed correctly. |
2 | Sample container definition in the CDS is not correct. | Check if the correct sample container is selected in the CDS. Verify if the dimension of the sample container match the database of your CDS. |
3 | Sample handler not aligned correctly. | Check if the sample handler can move freely. Do an autoreferencing (the needle assembly must be installed in the needle parkstation during this procedure). If this will not help: Please contact your Agilent service representative. |