Error ID: 25095
During the parking or movements of the needle assembly, the status information of the subparts is not read out successfully and the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | The sample loop capillary was squeezed in the needle parkstation. | Check if the sample loop is installed correctly. Do an autoreferencing (the needle assembly must be installed in the needle parkstation during this procedure). |
2 | The needle assembly was not installed correctly in the needle parkstation. | Check if the needle assembly is installed correctly. Install the needle assembly on the sample handler. Do a reset of the sample handler. Do an autoreferencing (the needle assembly must be installed in the needle parkstation during this procedure). If this will not help: Please contact your Agilent service representative. |
3 | Needle parkstation is loose. | Carefully tighten the needle parkstation. |
4 | For G5668A (Bio-inert Multisampler): Ceramic part of the needle is broken. | Exchange the needle assembly. |