Part Number (Description) | Rows | Columns | Plate height (mm) | Volume (μL) | Package | Closing mat compatibility |
---|---|---|---|---|---|---|
384Corning (No Agilent PN) | 16 | 24 | 14.4 | 80 |
|
|
384Nunc (No Agilent PN) | 16 | 24 | 14.4 | 80 |
|
|
5042-8502 (96-well plate, box of 25 sample plates) | 8 | 12 | 17.3 | 150 | 25 | 5042-1389 (Closing mat for all 96 Agilent plates) |
5065-4402 (96CappedAgilent) | 8 | 12 | 47.1 | 300 | 1 | 5042-1389 (Closing mat for all 96 Agilent plates) |
96Corning (No Agilent PN) | 8 | 12 | 14.3 | 300 |
|
|
96CorningV (No Agilent PN) | 8 | 12 | 14.3 | 300 |
|
|
96DeepNunc31mm (No Agilent PN) | 8 | 12 | 31,5 | 1000 |
|
|
96DeepRitter41mm (No Agilent PN) | 8 | 12 | 41.2 | 800 |
|
|
96Greiner (No Agilent PN) | 8 | 12 | 14.3 | 300 |
|
|
96GreinerV (No Agilent PN) | 8 | 12 | 14.3 | 250 |
|
|
96Nunc (No Agilent PN) | 8 | 12 | 14.3 | 400 |
|
|
5043-9300 (Well plate 96/2.2 mL, square wells, U shape, PP, 41 mm, 30/pk) | 8 | 12 | 41 | 2200 | 30 | 5043-9319 (Mat 96 wells, square, pierceable, silicone 50/pk) |
5043-9302 (Well plate 96/2.0 mL, round wells, U shape, PP, 45 mm, 30/pk) | 8 | 12 | 45.3 | 2000 | 30 | 5043-9317 (Mat 96 wells, round, pierceable, silicone 50/pk), 5043-9318(Mat 96 wells, round, piercable, silicone 100/pk) |
5043-9305 (Well plate 96/1.0 mL, round wells, U shape, PP, 32 mm, 50/pk) | 8 | 12 | 32 | 1000 | 50 | 5043-9317 (Mat 96 wells, round, pierceable, silicone 50/pk), 5043-9318 (Mat 96 wells, round, piercable, silicone 100/pk) |
5043-9308 (Well plate 96/1.2 mL, round wells, U shape, PP, 27 mm, 25/pk) | 8 | 12 | 27 | 1200 | 25 | 5043-9317 (Mat 96 wells, round, pierceable, silicone 50/pk), 5043-9318(Mat 96 wells, round, piercable, silicone 100/pk) |
5043-9309 (Well plate 96/1.2 mL, round wells, U shape, PP, 27 mm, 50/pk) | 8 | 12 | 27 | 1200 | 50 | 5043-9317 (Mat 96 wells, round, pierceable, silicone 50/pk), 5043-9318 (Mat 96 wells, round, piercable, silicone 100/pk) |
5043-9310 (Well plate 96/0.5 mL, round wells, U shape, PP, 14 mm, 30/pk) | 8 | 12 | 14 | 500 | 30 | 5043-9317 (Mat 96 wells, round, pierceable, silicone 50/pk), 5043-9318 (Mat 96 wells, round, piercable, silicone 100/pk) |
5043-9311 (Well plate 96/0.5 mL0.5 mL, round wells, U shape, PP, 14 mm, 120/pk | 8 | 12 | 14 | 500 | 120 | 5043-9317 (Mat 96 wells, round, pierceable, silicone 50/pk), 5043-9318 (Mat 96 wells, round, piercable, silicone 100/pk) |
5043-9312 (Well plate 96/0.33 mL, round wells, V shape, PP, 14 mm, 25/pk) | 8 | 12 | 14 | 330 | 25 | 5043-9317 (Mat 96 wells, round, pierceable, silicone 50/pk), 5043-9318 (Mat 96 wells, round, piercable, silicone 100/pk) |
5043-9313 (Well plate 96/0.33 mL, round wells, V shape, PP, 14 mm, 50/pk) | 8 | 12 | 14 | 330 | 50 | 5043-9317 (Mat 96 wells, round, pierceable, silicone 50/pk), 5043-9318 (Mat 96 wells, round, piercable, silicone 100/pk) |
5043-9314 (Well plate 96/0.33 mL, round wells, V shape, PP, 14 mm, 100/pk) | 8 | 12 | 14 | 330 | 100 | 5043-9317 (Mat 96 wells, round, pierceable, silicone 50/pk), 5043-9318 (Mat 96 wells, round, piercable, silicone 100/pk) |
5043-9315 (Well plate 384/0.19 mL, square wells, V shape, PP, 22 mm, 25/pk) | 16 | 24 | 22 | 190 | 25 | 5043-9320 (Mat 384 wells, square, pierceable, silicone 50/pk) |
Description (part number) | Rows | Columns | Plate height (mm) | Volume (µL) | Package |
---|---|---|---|---|---|
G2255-68700 (Vial plate for 54 x 2 mL vials (6/pk)) | 6 | 9 | 36 | 2000 | 6 |
5023-2471 (Vial plate 40 x 2 mL vials) | 5 | 8 | 43 | 2000 | 1 |
5022-6539 (Vial plate for 15 x 6 mL vials (1/pk)) only compatible with 3H drawers | 3 | 5 | 42 | 6000 | 1 |
5022-6538 (Vial plate for 27 Eppendorf tubes (1/pk)) | 3 | 9 | 40 | 500 – 2000 | 1 |
NOTE
For good chromatographic results the maximum filling should not exceed 3/4 of the total volume of the vial.
NOTE
Agilent Technologies recommends to use preslit septa.
NOTE
Bottom sensing is a feature to detect the depth of vials or plates via the software.
If the bottom sensing feature is used, the bottom of the plates and vials must resist the needle. Make sure that the material supports this feature. Inserts with flexible support should not be used.
NOTE
The default needle height offset value (0 mm) equates to an approximate distance of 2 mm above the bottom of a wellplate or a standard 2 mL vial at the reference bar, whereas it corresponds to an approximate distance of 5 mm above the bottom of a standard 2 mL vial in a vial tray. Using vial inserts or high recovery vials will impact the apparent distance between the needle tip and the bottom of the vessel.
NOTE
In case of custom-made wellplates or vials, please keep in mind the physical limitations of each drawer.
The maximum total height allowed (including sample container and vial caps, if present) is:
1H: 19 mm
2H: 45 mm
3H: 50 mm
NOTE
Adhesive foils are not recommended to seal wellplates. Alternatively, plates can be sealed with a 06644-001 (Pierceable aluminium foil) .