For bio-inert modules use bio-inert parts only! |
| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G4267-87012 | High Pressure Needle Seat, 0.12 mm (PEEK) OR |
1 | ![]() | G4267-87020 | High Pressure Seat Assembly 0.075 mm (PEEK) OR |
1 | ![]() | G5668-87017 | Bio-inert Seat ID 0.17 (for G5668A) |
1 | ![]() | G7137-87012 | High pressure seat assembly 0.12 mm Biocompatible (for G7137A) |
1 | ![]() | G7137-60075 | High Pressure Seat Bio ULD |
Prerequisites
|
WARNING
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
NOTE
Refer the Agilent 1290 Infinity III Ultra Low Dispersion Kit Technical Note (1290-UltraLowDispersion-TecPu-en-SD-29002050.pdf, SD-29002050) for further details.
In the Local Controller start the maintenance mode and select Change needle/seat function.
OR
In the Agilent Lab Advisor software select Service & Diagnostics in the system screen Maintenance Positions > Change Needle , click Start and wait until the needle assembly is in maintenance position.
Slightly pull (1.) the front clip which holds the needle seat in position. Then carefully lift up (2.) the complete leak tube needle assembly from the holder.
In the Local Controller close Change needle /seat.
OR
In the Agilent Lab Advisor software Change needle click End and wait until the needle assembly is in the needle park position.
Perform a Pressure Test.
NOTE
Always flush the sampler with fresh solvent and perform a pressure test before using the Auto Clean or Prime function after exchanging a needle seat.