Error ID: 22014
The pressure has exceeded the upper pressure limit.
Parameter: Measured pressure
Probable cause | Suggested actions | |
---|---|---|
1 | Blockage in flow path after the pressure sensor. | Check for blockages in the LC system, e.g. purge valve, Jet Weaver, degraded column, column frits, needle, needle seat, capillaries etc. Check for particles in the solvent. |
2 | Inappropriate settings (pressure limit, flow rate). | Decrease flow rate. Increase pressure limit. |
base-id: 3760556939
id: 27021601524779915