The communication board provides the 6-bit configuration switch and is located at the rear of the module. Switch settings provide configuration parameters for LAN and instrument specific initialization procedures.
All modules with communication board:
Default is ALL switches DOWN (best settings).
Default IP address for LAN 192.168.254.11
For specific LAN modes switches 4-5 must be set as required.
For boot resident/cold start modes switches 1+2 or 6 must be UP.
SW1 | SW2 | SW3 | SW4 | SW5 | SW6 | Mode | Init Mode |
---|---|---|---|---|---|---|---|
0 | 0 | 0 | 0 | 0 | 0 | COM | Use Default IP Address (192.168.254.11, Subnet mask: 255.255.255.0) |
0 | 0 | 0 | 0 | 1 | 0 | COM | Use Stored IP Address |
0 | 0 | 0 | 1 | 0 | 0 | COM | USE DHCP to request IP Address (Host name will be the MAC address) |
1 | 0 | 0 | 0 | 0 | 0 | Test | Boot Main System/Keep Data |
1 | 1 | 0 | 0 | 0 | 0 | Test | Boot Resident System/Keep Data |
1 | 0 | 0 | 0 | 0 | 1 | Test | Boot Main System/Revert to Default Data |
1 | 1 | 0 | 0 | 0 | 1 | Test | Boot Resident System/Revert to Default Data |
0 (switch down), 1 (switch up), SW (switch)
base-id: 3526771979
id: 27021601290994955