Error ID: 7453
The lamp heater current in the detector is missing. During lamp ignition, the processor monitors the heater current. If the current does not rise above the lower limit within 1 , the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Lamp disconnected. | Ensure the lamp connector is seated firmly. Ensure the lamp is connected. |
2 | Ignition started without the top foam in place. | Please contact your Agilent service representative. |
3 | Fan not running (permitting lamp on). | Please contact your Agilent service representative. |
4 | Defective mainboard. | Please contact your Agilent service representative. |
5 | Defective or non-Agilent lamp. | Exchange the lamp. |
6 | Defective UV lamp or non-Agilent lamp. | Exchange the UV lamp. |
7 | Defective power supply. | Please contact your Agilent service representative. |