Error ID: 7452
The lamp failed to ignite. The processor monitors the lamp current during the ignition cycle. If the lamp current does not rise above the lower limit within 2 – 5 s, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Lamp too hot. Hot gas discharge lamps may not ignite as easily as cold lamps. | Switch off the lamp and allow it to cool down for at least 15 minutes. |
2 | Lamp disconnected. | Ensure the lamp connector is seated firmly. Ensure the lamp is connected. |
3 | Defective or non-Agilent lamp. | Exchange the lamp. |
4 | Defective UV lamp or non-Agilent lamp. | Exchange the UV lamp. |
5 | Defective power supply. | Please contact your Agilent service representative. |
6 | Defective mainboard. | Please contact your Agilent service representative. |
base-id: 3284033803