Error ID: 7450
The lamp anode current is missing. The processor continually monitors the anode current drawn by the lamp during operation. If the anode current falls below the lower current limit, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Lamp disconnected. | Ensure the lamp connector is seated firmly. Ensure the lamp is connected. |
2 | Top foam removed while lamp is on. | Please contact your Agilent service representative. |
3 | Defective or non-Agilent lamp. | Exchange the lamp. |
4 | Defective UV lamp or non-Agilent lamp. | Exchange the UV lamp. |
5 | Defective mainboard. | Please contact your Agilent service representative. |
6 | Defective power supply. | Please contact your Agilent service representative. |