The following section provides useful information about Agilent capillaries and its characteristics.
Type - Material - Capillary dimensions - Fitting Left/Fitting right
Code provided with the part | Meaing of the code |
---|---|
Color code: | Material of the product is MP35N, the inner diameter is 0.20 or 0.25 mm |
Capillary | The part is a connection capillary |
MP35N | Material of the part is MP35N |
0.25 x 80 mm | The part has an inner diameter of 0.25 mm and a length of 80 mm |
SI/SI | Left fitting: Swagelok + 1.6 mm Port id, Intermediate Right fitting: Swagelok + 1.6 mm Port id, Intermediate |
To get an overview of the code in use, see
Type: Type (gives some indication on the primary function, like a loop or a connection capillary)
Material: Material (indicates which raw material is used for the capillary)
Dimension: Capillary dimensions (indicates inner diameter (id), length, and volume of the capillary)
Fittings: Fitting left/fitting right (indicates which fitting is used on both ends of the capillary)
Internal diameter in mm |
|
| Color code |
---|---|---|---|
0.015 |
|
| Orange |
0.025 |
|
| Yellow |
0.05 |
|
| Beige |
0.075 |
|
| Black |
0.075 | MP35N |
| Black with orange stripe |
0.1 |
|
| Purple |
0.12 |
|
| Red |
0.12 | MP35N |
| Red with orange stripe |
0.17 |
|
| Green |
0.17 | MP35N |
| Green with orange stripe |
0.20 /0.25 |
|
| Blue |
0.20 /0.25 | MP35N |
| Blue with orange stripe |
0.3 |
|
| Grey |
0.50 |
|
| Bone White |
NOTE
As you move to smaller-volume, high efficiency columns, you’ll want to use narrow id tubing, as opposed to the wider id tubing used for conventional HPLC instruments.
Key | Description |
---|---|
Capillary | Connection capillaries |
Loop | Loop capillaries |
Seat | Autosampler needle seats |
Tube | Tubing |
Heat exchanger | Heat exchanger |
Key | Description |
---|---|
ST | Stainless steel |
Ti | Titanium |
PK | PEEK |
FS/PK | PEEK-coated fused silicaFused silica in contact with solvent |
PK/ST | Stainless steel-coated PEEKStainless steel-coated PEEK |
PFFE | PTFE |
FS | Fused silica |
MP35N | Nickel-cobalt-chromium-molybdenium alloy |
[1]Fused silica in contact with solvent
[2]Stainless steel-coated PEEK
Description |
---|
id (mm) x Length (mm) |
Volume (µL) |
Key | Description |
---|---|
W | Swagelok + 0.8 mm Port id |
S | Swagelok + 1.6 mm Port id |
M | Metric M4 + 0.8 mm Port id |
E | Metric M3 + 1.6 mm Port id |
U | Swagelok union |
L | Long |
X | Extra long |
H | Long head |
G | Small head SW 4 |
N | Small head SW 5 |
F | Finger-tight |
V | 1200 bar |
B | Bio |
P | PEEK |
I | Intermediate |
base-id: 3003014411