The status indicator of the Sample Cooler/Sample Thermostat is incorporated in the graphical user interface (GUI) of the hosting sampler, which appears automatically when the unit is configured in the chromatography data system (CDS). When the cooler/thermostat is turned on, the set temperature and the actual temperature are also displayed.
where
1 | Cooler/Thermostat: Status indicator (Off) |
2 | Cooler/Thermostat: Status indicator (On) |
3 | Cooler/Thermostat: Actual temperature |
4 | Cooler/Thermostat: Set temperature |
NOTE
The actual temperature may deviate from the set temperature by up to 3 °C, depending on the temperature setting and ambient conditions.
NOTE
If the actual temperature differs by more than ± 2 °C from the set temperature, a yellow highlight is visible around the temperature reading. This, however, will not prevent the system from starting a new analysis, unless the Enable Analysis > Temperature within +/- 2 °C function is selected.