Universal Agilent Max-Light standard cartridge cell with 10 mm optical path length provides high sensitivity (noise: < ± 3 µAU) and low peak dispersion for 2.1 , 3 and 4.6 mm id columns.
Programmable slit from 1 to 8 nm provides incident light conditions for rapid optimization of sensitivity, linearity and spectral resolution.
Multiple wavelength and full spectral detection at high sampling rate of 240 Hz.
A reliable and robust peak integration process due through less baseline drift.
Full spectral detection for compound identification by spectral libraries or verification of the separation quality with peak purity analysis for fast LC. Simultaneous detection of up to 8 signals for increased sensitivity and selectivity.
Wide linear range (typically up to 2.5 AU) enables simultaneous quantification of primary compounds, by-products and impurities.
Upgrade option to HDR DAD solution provides 30x wider linear range for samples with widely different concentration levels.
Radio frequency identification (RFID) technology for flow cells and lamp improve data security and traceability.
Electronic temperature control (ETC) provides high baseline stability and practical sensitivity under fluctuating ambient temperature and humidity conditions.
base-id: 5701963019
id: 18014404211445003