WARNING
Absence of ground connection
The absence of ground connection can lead to electric shock or short circuit.
WARNING
Electrical shock hazard
Solvents may damage electrical cables.
WARNING
Heavy weight
The module is heavy.
WARNING
Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the power line at any time.
WARNING
Incorrect line voltage at the module
Shock hazard or damage of your instrument can result if the devices are connected to line voltage higher than specified.
WARNING
Hazard of electrical shock or damage of your instrumentation
can result, if the devices are connected to a line voltage higher than specified.
WARNING
Electrical shock hazard
The module is partially energized when switched off, as long as the power cord is plugged in.
The cover protects users from personal injuries, for example electrical shock.
WARNING
Unintended use of power cords
Using power cords for unintended purposes can lead to personal injury or damage of electronic equipment.
WARNING
Use of unsupplied cables
Using cables not supplied by Agilent Technologies can lead to damage of the electronic components or personal injury.
CAUTION
Condensation within the module
Condensation can damage the system electronics.
CAUTION
"Defective on arrival" problems
If there are signs of damage, please do not attempt to install the module. Inspection by Agilent is required to evaluate if the instrument is in good condition or damaged.
CAUTION
Solvent incompatibility
The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor. The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
NOTE
Agilent recommends that you install the HPLC instrument in the InfinityLab Flex Bench rack. This option helps to save bench space as all modules can be placed into one single stack. It also allows to easily relocate the instrument to another lab.