To protect optimal functionality of your flow cell:
G5615-60022 (Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar (12 MPa) for MWD/DAD, includes 0890-1763 – 0.18 x 1500 mm PEEK capillary and 5063-6591 – PEEK fittings) (PEEK, ceramic, sapphire, PTFE) for 1260 Infinity III Diode Array Detectors (G7115A):
The recommended pH range of the cell is 1 – 13 (short term 14)
G5615-60005 (Bio-inert flow cell, 8 µL, 20 bar) , (PEEK, fused silica, PTFE) for 1260 Infinity III Fluorescence Detector (G7121A/B)
The recommended pH range of the cell is 1 – 12 (solvent dependent).
If the flow cell is transported while temperatures are below 5 °C, it must be ensured that the cell is filled with alcohol to avoid damage by freezing water.
Aqueous solvents in the flow cell can build up algae. Therefore, do not leave aqueous solvents sitting in the flow cell. Add a small percentage of organic solvents (for example, about 5 % of acetonitrile or methanol).