Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7120-68708 | InfinityLab System Tool kit |
In Lab Advisor go to Service & Diagnostics > Remove/Install Pump Head and follow instructions given on the screen.
DO NOT REMOVE the heat exchanger connection between the pump heads marked by the red X.
Turn the inlet valve 90 ° counterclockwise (1) and remove the inlet weaver from the inlet valve (2).
Loosen the inlet valve but keep it installed.
Loosen the outlet valve but keep it installed.
Open the screws step by step, not screw by screw.
Remove the complete pump head assembly by holding both heads and pulling it to the front.
base-id: 4133707659