When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]()
|
| Hexagonal key, 4 mm |
1 | ![]() | 01018-23702 | Insert tool |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5063-6589 | Metering seal (pack of 2) for 100 µL analytical head , OR |
1 | ![]() | 0905-1420 | PE seal (pack of 2) |
1 | ![]() | 5063-6586 | Sapphire piston, 100 µL |
Prerequisites
|
NOTE
Both pump head assemblies use the same internal components. In addition, pump head A is fitted with the purge valve. The following procedure describes the removal and disassembly of pump head A (left). For pump head B (right) proceed in the same way and skip steps that deal with the purge valve.
Install the pump head, see Reinstall the Pump Head Assembly.
base-id: 3836362379
id: 18014402345844363