When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 8710-1924 | Open-end wrench 14 mm |
1 | ![]()
|
| Pair of tweezers , OR |
1 | ![]()
|
| Toothpick |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 01018-22707 | PTFE Frit (5/Pk) |
1 | ![]() | G7111-60061 | Purge valve |
1 | ![]() | 5067-4728 | Seal cap assembly |
Prerequisites
|
Using a pair of tweezers or a toothpick remove the frit.
Reinstall the seal cap including the gold seal.
NOTE
Before reinstallation always check the gold seal in the seal cap. A deformed seal cap should be exchanged.
Run the System Pressure Test (see System Pressure Test).
base-id: 3836343819