When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5023-2524 | Hex Key Set |
1 | ![]() | 5067-5688 | Torque wrench 1 - 25 Nm with 14 mm wrench |
1 | ![]() | G4220-20013 | 4 mm hex bit |
1 | ![]() | G4220-20015 | Adapter ¼ in square to hex |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 79846-65501 | PTFE lubricant |
CAUTION
Damage of the pump drive
Starting the pump when the pump head is removed may damage the pump drive.
If needed, apply a small amount of grease on the back of the screws. Normally, the grease added during manufacturing is sufficient for a long time.
If installed, reconnect the active inlet valve cable to its connector.
If a standard seal has been installed, run the seal wear-in procedure, which includes a replacement of the purge valve frit.
For the normal phase seal, the purge valve frit should be replaced, see Exchange the Purge Valve Frit or the Purge Valve.
Run the Leak Rate Test (see Leak Rate Test).
base-id: 3821925771