It is necessary to perform periodic inspection of the instrument to ensure its safe use. It is possible to have these periodic inspections performed by Agilent service representatives on a contractual basis. For information regarding the maintenance inspection contract, contact your Agilent representative.
The following pages describe maintenance (simple repairs) of the pump that can be carried out without opening the main cover.
Procedure | Typical Frequency | Notes |
---|---|---|
If solvent filter is blocked |
Gradient performance problems, intermittent pressure fluctuations | |
If internally leaking |
Pressure ripple unstable, run Leak Rate Test for verification | |
If internally leaking |
Pressure ripple unstable, run Leak Rate Test for verification | |
If internally leaking |
Pressure ripple unstable, run Leak Rate Test for verification | |
If internally leaking |
Solvent dripping out of waste outlet when valve closed | |
If the frit shows indication of contamination or blockage |
A pressure drop of > 10 bar across the frit (at a water flow of 5 mL/min with open purge valve) indicates blockage | |
If pump performance indicates seal wear |
Leaks at lower pump head side, unstable retention times, pressure ripple unstable - run Leak Rate Test for verification | |
Exchange pistons, see Maintenance of a Pump Head Without Seal Wash Option |
If scratched |
Seal life time shorter than usual - check pistons while changing the seals |