Error ID: 2015, 2501
The system pressure has fallen below the lower pressure limit.
Probable cause | Suggested actions | |
---|---|---|
1 | Solvent bottle empty. | Replenish solvent. |
2 | Lower pressure limit set too high. | Ensure the lower pressure limit is set to a value suitable for the analysis. |
3 | Air bubbles in the mobile phase. | Make sure that the degasser is in flow path and works correctly. Purge the module. Ensure solvent inlet filters are not blocked. |
4 | Air bubbles in the mobile phase. | Ensure bubble free flow. Purge the preparative pump. |
5 | Leak. | Inspect the pump head, capillaries and fittings for signs of a leak. Purge the module. Run a pressure test to determine whether the seals or other module components are defective. |
6 | Defective damper. | Please contact your Agilent service representative. |
7 | Defective pressure sensor. | Please contact your Agilent service representative. |
8 | Defective mainboard. | Please contact your Agilent service representative. |
9 | Defective active inlet valve (AIV). | Replace AIV cartridge. |
10 | Defective outlet ball valve (OBV). | Replace OBV. |