Error ID: 2055, 2524
Pump firmware version A.02.32 and higher allow to set solvent bottle fillings in the data system. If the volume level in the bottle falls below the specified value the error message appears when the feature is configured accordingly.
The error message is triggered if the remaining volume in a solvent bottle falls below the set limit.
Probable cause | Suggested actions | |
---|---|---|
1 | Volume in bottle below specified volume. | Refill bottles and reset solvent counters. |
2 | Incorrect setting. | Make sure the limits are set correctly. Make sure the set solvent volume matches the actual bottle filling and set the shutoff limit to a reasonable value (e.g. 100 mL for 1 L bottles). |