Each movement of the sampling sequence can be done under manual control. This is useful during troubleshooting, where close observation of each of the sampling steps is required to confirm a specific failure mode or verify successful completion of a repair. Each injector step command actually consists of a series of individual commands that move the multisampler components to predefined positions, enabling the specific step to be done.
Run the Injector Steps in the Service & Diagnostics View in the Agilent Lab Advisor (for further information see Online-Help of user interface).
Select the individual step command like needle selection and needle position (for further information see Online-Help of user interface).
NOTE
Follow a logical order to use the injector steps function.