The sampler auto referencing uses predefined positions on the base plate and the sample hotel to calibrate the positioning of the needle parkstation and the sample hotel. The auto referencing is required to compensate deviations in positioning the needle assembly and the sample tray. The auto referencing is required after disassembling the system or when you exchange the sample handler, the sample hotel, the needle parkstation, the needle assembly or one of the mainboards. This function is implemented in the drawer detection and in the needle exchange routine.
When
|
Prerequisites
|
Open the Agilent chromatography data system of the instrument.
A right-click into the Active Area of the sampler will open a menu to modify
drawer configuration
capillaries
Reference Vial Bar
Use drawer configuration and follow the software instructions.
Auto referencing is done.
Click the Back button to leave the Service & Diagnosis menu.
NOTE
For auto referencing, you can alternatively use the Local Controller.