The test determines the leak rate of the system between pump outlet valves and a blank nut. The blank nut can be positioned at different locations in the system before the flow cell, to determine and verify the leak rate of individual modules and components. The test allows for setting the pressure at which the test is performed. The leak rate of high pressure parts is not always a linear function and therefore it is recommended to perform the test at a pressure that corresponds to the normal operating pressure of the system.
When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5067-6127 | Blank nut SL For the 1290 Infinity III Multisampler, use the Blank Nut SL which fits the special port size of the VICI valve. This Blank Nut is backward compatible and can be used for the 1260 Infinity III Multisampler as well. OR |
1 | ![]() | 5043-0277 | PEEK blank nut for bio-compatible devices For Bio-inert or 1290 Bio systems |
Run the System pressure test with the Agilent Lab Advisor (for further information see Online-Help of user interface).