For best performance of the Multisampler
When using the Multisampler in a system with a vacuum degassing unit, shortly degas your samples before using them in the Multisampler.
NOTE
To achieve good chromatographic results, it is mandatory to use degassed solvents with the sampler.
Filter samples before use in an InfinityLab LC Series system. Use 5067-6189 (InfinityLab Inline Filter (0.3 µm)) for inline filtering.
NOTE
This inline filter contains stainless steel and is not indicated for use in bio-inert or biocompatible systems.
When using buffer solutions, flush the system with water before switching it off.
Check the Multisampler plungers for scratches, grooves, and dents when changing the piston seal. Damaged plungers cause micro leaks and will decrease the lifetime of the seal.
Solvent Information: Observe recommendations on the use of solvents, see Solvent Information.