Error ID: 30822
The heating function of the Integrated Column Compartment is switched off due to the malfunctioning of the digital temperature sensor.
To verify the correct functioning of the heating element, perform the Heater Test in the Agilent Lab Advisor software.
Probable cause | Suggested actions | |
---|---|---|
1 | Connection problems. | Power off the system and verify that the ICC is correctly mounted in the sampler. Power on the system and check if the problem persists. |
2 | Potential hardware error. | Please contact your Agilent service representative. |