Error ID: 30824
The heating function of the Integrated Column Compartment is switched off because the temperature registered by the temperature sensor exceeds the upper limit (100 °C).
To verify the correct functioning of the heating element, perform the Heater Test in the Agilent Lab Advisor software
Probable cause | Suggested actions | |
---|---|---|
1 | Connection problems. | Power off the system and verify that the ICC is correctly mounted in the sampler. Power on the system and check if the problem persists. |
2 | Potential hardware error. | Please contact your Agilent service representative. |