Error ID: 30820
The system reported an error because reaching the set temperature took longer than expected. To verify the correct functioning of the heating, perform the Heater Test in the Agilent Lab Advisor software.
Probable cause | Suggested actions | |
---|---|---|
1 | Connection problems. | Power off the system and verify that the ICC is correctly mounted in the sampler. Power on the system and check if the problem persists. |
2 | Potential hardware error. | Please contact your Agilent service representative. |
base-id: 3716051083
id: 18014402225533067