Error ID: 30715
The cooler/thermostat was switched off due to a condensate event.
Probable cause | Suggested actions | |
---|---|---|
1 | Overfilled container. | Empty the condensate container. Verify that the open end of the tubing doesn’t immerse in the liquid. |
2 | Drainage issues. | Verify the correct plumbing of the condensate drainage system. Make sure that no kinks or mechanical blocks are present in the drainage system. Avoid the formation of the siphoning effect. Make sure that the hosting sampler is level. |