Error ID: 34319
The injection valve position is not found or the valve is unable to switch.
Probable cause | Suggested actions | |
---|---|---|
1 | Overtightened screws. | Verify that the stator screws are not overtightened as it can lead to hindered rotor movements. If necessary, loosen the stator screws and retighten them appropriately. |
2 | Broken rotor seal. | Check the rotor seal for signs of visible damage and if necessary, replace it. |
3 | Broken stator. | Check the stator for signs of visible damage and if necessary, replace it. |
4 | Potential hardware error. | Please contact your Agilent service representative. |