Error ID: 34303, 34304
Error ID: 34303 → Needle up failed
Error ID: 34304 → Needle down failed
The upper or the bottom position of the needle arm is not found or the needle arm is unable to move.
Probable cause | Suggested actions | |
---|---|---|
1 | Physical blockage. | Look for potential sources of physical blockage in the path of the needle arm and eliminate them. Verify the correct functioning of the needle arm by using the Injector Steps Lab Advisor tool. Restart the sampler and observe if the problem persists. |
2 | Potential hardware error. | Please contact your Agilent service representative. |