With the Sample Thermostat installed, the Control dialog box of the hosting sampler will include the following thermostat-specific control options:
At Power On:
Turn On Thermostat: The thermostat turns on automatically upon powering on the sampler.
Thermostat:
On: The thermostat turns on and the system starts to regulate the temperature inside the sample space towards the setpoint.
Off: The thermostat turns off.
Enable Analysis
NOTE
The Enable Analysis control setting is available since LC & CE drivers A.02.19.
With any temperature: The analysis starts regardless of the actual temperature inside the sampler.
Temperature within +/- 2 °C: The analysis starts only when the actual temperature is within the ± 2 °C range of the setpoint temperature.
NOTE
The Temperature within +/- 2 °C option is only available for the Sample Thermostat.