WARNING
Flammable refrigerant
Formation of flammable gas-air mixtures inside the Sample
Thermostat and laboratory.
- Keep open fire or
sources of ignition away from the device.
- Ensure a
room size of 4 m3 (1 m3
for every 8 g of R600a refrigerant
inside of the Sample Thermostat).
- Ensure adequate
ventilation: typical air exchange of 25
m3/h per m² of laboratory
floor area.
- Keep all ventilation openings in the
enclosure clear of obstructions. Do not block the openings on the circumference
of the Sample Thermostat.