Error ID: 1034
This error message comes up when the communication between the optical unit and the main board has a problem.
Probable cause | Suggested actions | |
---|---|---|
1 | Random communication error. | Switch the detector off and on again at the on/off switch. If the error reoccurs: Please contact your Agilent service representative. |
2 | Defective detector main board. | Please contact your Agilent service representative. |
3 | Defective PDA/optical unit. | Please contact your Agilent service representative. |
base-id: 3359355403