Error ID: 1018
The visible lamp current is missing.
The processor continually monitors the lamp current during operation. If the current falls below the lower current limit, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | Lamp disconnected. | Ensure the lamp connector is seated firmly. Ensure the lamp is connected. |
2 | Defective visible lamp. | Exchange the visible lamp. |
3 | Defective connector or cable. | Please contact your Agilent service representative. |
4 | Defective power supply. | Please contact your Agilent service representative. |