| For bio-inert modules use bio-inert parts only! Do not mix with bio / biocompatible parts. |
| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
CAUTION
The valve actuator contains sensitive optical parts, which need to be protected from dust and other pollution. Pollution of these parts can impair the accurate selection of valve ports and therefore bias measurement results.
CAUTION
Column Damage or Bias Measurement Results
Switching the valve to a wrong position can damage the column or bias measurement results.
CAUTION
Valve Damage
Using a low pressure valve on the high pressure side can damage the valve.
NOTE
For information about the compatibility mode of 800 bar valve heads see 01200-90134 (Information on RFID Tag Technical Note) .
NOTE
For a correct installation of the valve head, the outside pin (red) must completely fit into the outside groove on the valve drive’s shaft (red). A correct installation is only possible if the two pins (green and blue) on the valve head fit into their corresponding grooves on the valve drive’s actuator axis. Their match depends on the diameter of the pin and groove.
NOTE
The tag reader reads the valve head properties from the valve head RFID tag during initialization of the module. Valve properties will not be updated, if the valve head is replaced while the module is on. Selection of valve port positions can fail, if the instrument does not know the properties of the installed valve.
NOTE
To allow correct valve identification, power off the valve drive for at least 10 s.
NOTE
For firmware requirements see 01200-90134 (Information on RFID Tag Technical Note) which is included to each valve head.
CAUTION
Sample degradation and contamination of the instrument
Metal parts in the flow path can interact with the bio-molecules in the sample leading to sample degradation and contamination.
Insert the valve head into the valve shaft.
OR
If the outside pin does not fit into the outside groove, you have to turn the valve head until you feel that the two pins snap into the grooves. Now you should feel additional resistance from the valve drive while continuously turning the valve head until the pin fits into the groove.
When the outer pin is locked into the groove, manually screw the nut onto the valve head.
NOTE
Fasten the nut manually. Do not use any tools.
Install capillaries.
Push the valve head until it snaps in and stays in the rear position.
Power on or power-cycle your module, so the valve head gets recognized during module initialization.
base-id: 3001839499