| For bio-inert modules use bio-inert parts only! Do not mix with bio / biocompatible parts. |
| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
This procedure describes the exchange of a column for a different one.
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5043-0915 | Fitting mounting tool, OR |
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
For 1290
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5067-5957 | InfinityLab Quick Connect Assy ST 0.12 mm x 105 mm (from heat exchanger to column inlet) |
1 | ![]() | 5500-1191 | InfinityLab Quick Turn Capillary ST 0.12 mm x 280 mm, long socket (from column to detector) |
1 | ![]() | 5067-5966 | InfinityLab Quick Turn Fitting |
Parts required
For 1260
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5500-1193 | InfinityLab Quick Turn Capillary ST 0.17 mm x 105 mm, long socket (from heat exchanger to column) |
1 | ![]() | 5500-1191 | InfinityLab Quick Turn Capillary ST 0.12 mm x 280 mm, long socket (from column to detector) |
1 | ![]() | 5067-5966 | InfinityLab Quick Turn Fitting (column inlet) |
1 | ![]() | 0100-1516 | Finger-tight fitting PEEK, 2/pk (column outlet) |
Parts required
Bio-inert option
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5067-4780 | Bio Capillary 0.17 mm x 300 mm, 1.6 mm OD socket (from column to detector) |
1 | ![]() | 5067-5966 | InfinityLab Quick Turn Fitting (column inlet) |
1 | ![]() | 0100-1516 | Finger-tight fitting PEEK, 2/pk (column outlet) |
Parts required
Bio-compatible option
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5067-5966 | InfinityLab Quick Turn Fitting |
1 | ![]() | 5500-1596 | Quick Turn Capillary MP35N 0.12 mm x 280 mm |
1 | ![]() | 5500-1578 | Quick Connect Capillary MP35N 0.12 mm x 105 mm |
1 | ![]() | 5067-5965 | InfinityLab Quick Connect LC fitting |
Prerequisites
Remove the column from the column holder.
Disconnect the column outlet.
Open the lever of the Quick Connect Fitting.
Unscrew and remove the Quick Connect Fitting from the column inlet.
Connect the column inlet capillary to the column inlet using the Quick Connect Fitting.
Turn the column onto the Quick Connect Fitting until it is finger-tight.
Close the lever to make a leak-tight connection.
Connect the capillary to detector to the column outlet. When using the Quick Turn Fitting, tighten the fitting until it is finger-tight.
Use a wrench to counter the column while tightening the capillary fitting.
Position the column into the column holder clip(s).
Exchange old column with new column of the same type: no further action required.
OR:Exchange old column with new column of different type: --- MISSING LINK ---.