Test failed
NOTE
Notice the difference between error in the test and a failed result! An error is caused by an abnormal termination during the operation of the test, whereas a failed result indicates that the test result were not within the specified limits.
Probable cause | Suggested actions | |
---|---|---|
1 | Solvent line and pump is not purged properly. | Repeat the pump purge. |
2 | Solvent filter is blocked. | Replace the solvent filter. |
3 | Bottle plates is not calibrated. | Perform bottle plates calibration. |
4 | Module electronic: Sensing modules or Mainboard is defect. | Contact your local Agilent representative. |
base-id: 12232810251