During an analysis, the internal synchronization or communication between one or more of the modules in the system has failed. The system processors continually monitor the system configuration. If one or more of the modules is no longer recognized as being connected to the system, the error message is generated.
Probable cause | Suggested actions | |
---|---|---|
1 | CAN cable disconnected. | Ensure all the CAN cables are connected correctly. |
2 | Defective CAN cable. | Switch off the system. Restart the system, and determine which module or modules are not recognized by the system. |
3 | Defective valve. | Exchange the defective Preparative Valve-Based Fraction Collector. |
4 | Defective MTP main board in other module. | Exchange the potentially defective main boards. |
base-id: 11470229643
id: 11470229643