Type | Specification | Comments |
---|---|---|
Weight | 2.2 kg (4.9 lbs) |
|
Dimensions | 95 x 95 x 300 mm (3.7 x 3.7 x 11.8 inches) |
|
Line Voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line Frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 20 VA, 4 W |
|
Ambient operating temperature | 4–40 °C (39–104 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Installation category II, Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comments |
---|---|---|
External Leak Sensor | Yes | controlled via 2.5 mm jack |
Flow rate | 4 – 200 mL/min |
|
Maximum Pressure | 69 bar (100 psi) |
|
Valve Options | 5067-4292 Quick-Change valve head (user-exchangeable) Valve POD Rotor/Stator seals: 2000000 actions (switches), 1000000 transfers |
|
Valve Identification | Via RFID tag |
|
Instrument Control | LC & CE Drivers A.02.17 or above Instrument Control Framework (ICF) A.02.04 or above InfinityLab LC Companion (G7108A) Lab Advisor B.02.10 or above | For details about supported software versions refer to the compatibility matrix of your version of the LC and CE Drivers |
Communications | CAN (2x) | G7170B is a hosted module. The LC stack needs to contain a LAN card for communication and control. |
Maintenance and safety-related features | Leak handling with external leak sensor and leak pane. Leak detection, safe leak handling, leak output signal for shutdown of pumping system, and low voltages in major maintenance areas Extensive diagnostics, error detection and display with Agilent Lab Advisor software |
|
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits and feedback messages. Electronic records of maintenance and errors |
|
Housing | All materials recyclable |
|