The Block Instrument for Service dialog box allows you to block the instrument.
Prerequisites
|
Navigate to Health >
Block for Service
Select Block Now.
In the Block Instrument for Service dialog, enter a title (mandatory). It will be displayed in the block banner and in related notifications.
Select Block Instrument.
The block has the following effect:
Chromatography Data Systems (CDS) with LC/CE Drivers version 3.10 or above are disconnected. If a CDS with LC/CE Drivers version 3.10 or above tries to connect or reconnect, a connection error is generated.
The Instrument Blocked for Service banner is displayed below the status bar on all screens in all sessions while the block is in place. It contains a link to view the Block for Service screen.
Users without the permission Activate/deactivate service mode cannot do the following:
Execute quick actions.
Start tasks.
Execute maintenance, diagnostic, or troubleshooting procedures.
Upload and install new versions of the Assist Control Software and module-specific firmware.
Modify security settings.
Export or import settings.
Create or restore backups.
Scheduled tasks that become active are immediately skipped.
Select Unblock on the Block for Service screen.
OR
Select Unblock in the Instrument Blocked For Service notification.
base-id: 13128835723
id: 9007212383576715