Type | Specification | Comments |
|---|---|---|
Weight | 15.9 kg (35.05 lbs) | Without flow cell |
Dimensions | 203 x 396 x 436 mm |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 25 W (idle) 240 VA; 240 W | Idle Lamp on + Multi-WL mode |
Ambient operating temperature | 4–40 °C (39–104 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: | Overvoltage category II, | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comments |
|---|---|---|
Detection type | Fluorescence Detector |
|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, |
|
Single wavelength operation | Raman SNR > 2500 Raman SNR vs Dark > 30000 | Raman SNR: Raman SNR vs Dark: response time = 4.2 sec PMT = Standard |
Light source | Hg-Xe arc lamp 150 W Lifetime 3000 h | Guaranteed lifetime = 1400 h, permissible 3000 h. Lamp lifetime based on intensity and operational safety. |
Maximum data rate | 160 Hz |
|
Wavelength range (Ex, Em) | Settable range 200 nm – 1200 nm and zero-order |
|
Spectral bandwidth (Ex, Em) | <= 20 nm |
|
Wavelength accuracy | +/- 3.0 nm |
|
Wavelength precision | +/- 0.2 nm |
|
Reference system | Reference diode |
|
Timetable programming | Wavelength (Ex, Em) PMT Gain |
|
Multi-channel acquisition | Up to 5 channels | Maximum data rate 5 Hz |
Spectrum acquisition | Excitation and Emission spectra | Scannable range 200 nm – 800 nm Step size 1 nm - 20 nm |
Flow cell volume | 2 µL volume 13 µL volume |
|
Maximum pressure | 60 bar (6 MPa) |
|
pH range | 1.0 ... 12.5 | Solvent dependent |
Wetted materials | MP35N, Fused Silica, FEP | Bio compatible |
Analog output | 100 mV or 1 V full scale | BNC connector type |
Instrument Control | LC & CE Drivers 3.11 or above Instrument Control Framework (ICF) 3.5 or above Lab Advisor 2.23 or above Assist Control Software 2.1 or above Instant Pilot (G4208A) with firmware B.02.26 or above | For details about supported software versions refer to the compatibility matrix of your version of the LC and CE Drivers |
Communication | LAN, CAN, USB, |
|
Safety features and maintenance | Leak detection, safe leak handling, leak output signal for shutdown of the pumping system. No hazardous voltages in major maintenance areas. Extensive diagnostics, error detection and display. |
|
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with pre-defined and user settable limits and feedback messages. Electronic records of maintenance and errors. |
|
Maintenance | Tool-free lamp exchange Tool-free flow cell exchange Tool-free lamp house window exchange | Tasks designed for safe customer handling |
Housing | All materials are recyclable. |
|
base-id: 12746848395