Error ID:
Probable cause | Suggested actions | |
|---|---|---|
1 | Data acquisition leads not connected |
Ensure connectors to computer or integrator are securely connected. |
2 | Light source inactive |
Check LED or LASER intensity is 100 %.
Check LED or LASER is functioning correctly, by stopping solvent flow, cycling the power. Then reading the offset value in RUN mode should be below 130 mV. |
3 | ProbableCause Output below 0 mV |
Stop pump flow and A/Z without liquid flowing. |
4 | ProbableCause Instrument in STANDBY mode |
Select RUN mode. |
5 | Nebuliser or nebuliser inlet tube blocked Nebulizer or nebulizer inlet tube blocked |
Manually syringe water into ELSD front Inlet port to remove obstruction. |