Error ID:
Probable cause | Suggested actions | |
|---|---|---|
1 | Inconsistent nebulization |
Nitrogen is the recommended gas - others can be used but may not nebulize as efficiently. |
2 | Incorrect gas being used |
Change gas to nitrogen or evaluate different nitrogen sources. |
3 | Poor regulation of inlet gases |
If using bottled gas, check that gas regulator is functioning correctly and giving consistent flow.
Alternatively, use a pulse dampener. |
4 | Insufficient smoothing |
The broader the peaks, the higher the smoothing value is required. Increase smoothing to 50 for Flash separations. |
5 | Sample precipitation during nebulization |
Reduce sample concentration or inject same loading in larger volume.
Check sample solubility in mobile phase eluents. |
6 | Inconsistent pump flow rates |
See remedy for pump pulsation in the baseline noise section. |