Error ID:
Probable cause | Suggested actions | |
|---|---|---|
1 | Ambient lab temperature too high |
Move detector to laboratory where ambient temperature is <25 °C. |
2 | Faulty Peltier cooler unit |
Please contact your Agilent service representative. |
base-id: 4344265611
id: 9007203599006603