Error ID:
Probable cause | Suggested actions | |
|---|---|---|
1 | Inefficient evaporation |
Increase the evaporator temperature and/or gas flow. |
2 | High concentration of non-volatile buffer or stabiliser High concentration of nonvolatile buffer or stabilizer |
Use a lower concentration of stabilizer, unstabilized solvent or, a more volatile buffer (ammonium acetate or ammonium formate). |
3 | Contaminated diffuser |
Perform cleaning procedure. |
4 | Optics Heater failed |
Please contact your Agilent service representative. |