Error ID:
Probable cause | Suggested actions | |
|---|---|---|
1 | ProbableCause Particulate matter in the gas supply |
Filter the incoming gas, or change the supply. |
2 | ProbableCause Column shedding |
Replace column or fit an inline filter with a 0.2 µm membrane filter directly after the column. |
3 | ProbableCause Poor nebulization |
Check solvent flow rate into ELSD is constant.
Check inlet gas flow is >60 psi and stable. |
4 | Insufficient evaporation |
Increase the temperature of the evaporator by 10 °C until the baseline noise decreases. |
5 | Non-volatile additive in mobile phase Nonvolatile additive in mobile phase |
Use a volatile mobile phase. |
base-id: 4344078219
id: 9007203598819211