Error ID: 29240
Reading the multipurpose valve tag failed.
Parameter: None
Probable cause | Suggested actions | |
|---|---|---|
1 | Reading of multi purpose valve tag failed Reading of multipurpose valve tag failed. | Check cable connection. |
2 | Multipurpose valve head taag defect or empty Multipurpose valve head tag defect or empty. | Replace multipurpose valve head. |
3 | Multipurpose valve tag reader is defect Multipurpose valve tag reader is defect. | Please contact your Agilent service representative. |
base-id: 4132814475
id: 9007203387555467