Dark-Current Test evaluation
Probable cause | Suggested actions | |
|---|---|---|
1 | Defective slit assembly (stray light) Defective slit assembly (stray light). |
Run the Self-Test. Run the Self-Test.
Run the Slit Test (G7117B) (part of the Self-Test). |
2 | Defective detector main board Defective detector main board. | Please contact your Agilent service representative. |
3 | Defective optical unit Defective PDA/optical unit. | Please contact your Agilent service representative. |