Error ID:
Holmium Oxide Test evaluation
Probable cause | Suggested actions | |
|---|---|---|
1 | Absorbing solvent or air bubble in flow cell Absorbing solvent or air bubble in flow cell. | Ensure the flow cell is filled with water, and free from air bubbles. |
2 | Incorrect calibration |
Recalibrate (see Wavelength Calibration) and repeat the test. |
3 | Dirty or contaminated flow cell Dirty or contaminated flow cell. |
Run the cell test (see Cell Test). If the test fails, exchange the flow cell windows. |
4 | Dirty or contaminated optical components Dirty or contaminated optical components (achromat, windows). |
Clean optical components with alcohol and lint-free cloth or replace the parts (see Intensity Test). |
5 | Old or non-Agilent lamp Old or non-Agilent lamp. | Exchange the UV lamp. |